Symptom

Less image space available than normal. See long description for examples. Less image space available than normal, even after removing all exams in Image Works. Example: H3 or GOC1 console should have 20,000 (512 images) available on a clean system with an Octane cpu with two 9 GB disk drives. But with the replacement of the 9GB drives with 18 and 73GB drives, image space may increase up to 130,000 (512 images) depending on software. Example: GOC2 console should have 100,000 (512 images) available on a clean system with an HP-8000 cpu with one 18GB disk and one 73GB disk. Example: GOC3 or 4 console should have 200,000 (512 images) available on a clean system with an HP-8200 cpu with one 36GB disk and two 73GB disks. Data above is approximate, but can be used to determine if there is a problem with disk space.

Error message

Error: 210000441 Image space allocation failed. See LD for complete error Mon Mar 26 12:51:14 2007 Error: 210000441 Host: roct Process: dms File: Initialize.cxx Line: 856 Image space allocation failed. size 3863552 Error code: -1

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5117866-5 73 GB HDD used in HPxw8200 GRE. See Long Discription for Details. Part #: 5117866-5 Part Type: GP CONSUMABLES Part Weight: -139998.60 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB HDD for Lighting System Controller 5117866-2 FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution