Pular para o conteúdo principal
GE Healthcare · CTLightSpeed210000434

LightSpeed 4.X - GRE

Symptom

After setting up scan, the customer goes to press the "Confirm" button it was not lit up. After setting up scan, the customer goes to press the "Confirm" button it was not lit up

Error message

Error: 2106 Image Storage Full See Long Description for complete error Wed Aug 30 16:07:53 2006 Error: 2106 Host: IRIS Process: imserver File: imsprim.c Line: 742 Image Storage Full

Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data See Long Description for complete error Fri Apr 21 07:41:23 2006 Error: 210000434 Host: lightspeed Process: dms File: SaveContext.cxx Line: 184 I MAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.