GE Healthcare · CTLightSpeed260115194
LightSpeed 5.X (60kW)
Symptom
Table will not drive up or down
Error message
Error: 260115194 Elevation drive fatal fault occurred. See long description Tue Nov 3 08:04:58 2009 Error: 260115194 Host: tgp Process: tgp_subsys File: mtcb_emc_action.c Line: 206 Tue Nov 3 08:04:57 2009 Host : Table Sub-System Ermes # : 260115194 Exception Class : Secondary Severity : Pri/Soft File : mtcb_emc_action.c Line# : 206 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation drive fatal fault occurred. FaultCode : Not Assigned ExtFault : Not Assigned Firmware can not reset the fault any longer. Please check followings then try to reset the MTCB board. + Cable connection around the driver.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2342729-3 Elevation or Cradle Drive Amplifier used on the 5.X table - same driver amp is used for both See LD for additional part information Part #: 2342729-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER TABLE SERVO DRIVE, EMC EDITION 2 COMPLIANT FRU: Yes Status: Orderable Comments: **Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money.** Part added per DST-HP60 16 slice RSPL. 040906 MS13452.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.