Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26011813826011828326011822650044

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118138 LVPS400:gate + 24V too low Mon Jul 11 17:11:48 2011 Host : Jedi Ermes # : 260118138 Exception Class : Secondary Severity : Sec/Soft File : gen_exec.cxx Line# : 2616 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 (25-0526)LVPS400:gate + 24V too low. Z6,Z7= 4h,0h

Error: 260118283 (25-0593) LVPS 400: Power Suppy output set to off. see long description Tue Jul 1 17:20:59 2008 Error: 260118283 Host: orp Process: orp_subsys File: gen_exec.c Line: 2414 Tue Jul 1 17:20:58 2008 Host : Jedi Ermes # : 260118283 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 2414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 (25-0593)LVPS 400: Power Suppy output set to off. 3 over/under voltage FW errors in the same LVPS power supply output. Ext15V=0, Ext160V=1, Can15V=2, Gate24V=3, Main_inv=4, Heater1=5, Heater2=6,

Error: 260118226 Inverter Gate Power Supply error See the long description for the full error code Sat Dec 15 20:17:54 2007 Error: 260118226 Host: orp Process: orp subsys File: gen exec.c Line: 2414 Sat Dec 15 20:17:53 2007 Host : Jedi Ermes # : 260118226 Exception Class : Abort Severity : Prima

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
2281950-3 or 2281950-4 : JH4 INVERTER WITH SHIELDS used on LS 5.X (80/100), LS 7.X See Long Description for Details 2281950-3 Can also use 2281950-4 JH4 INVERTER WITH SHIELDS used on LS 5.X (80/100), LS 7.X Part #: 2281950- 3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2281950 and 2281950-2. ms13452 040923 Set as Critical Part for Product CT_VCT, ms04278, 05Oct14 - If you receive an inverter as FRU (P/n 2281950, 2281950-2, 2281950-3) that must be installed to replace a defective inverter on site, check first Part number of PPC kV control board in this FRU: - if it is labelled 2351200-11 rev 001 this board needs to be replaced with the kV control board that is installed in the defective inverter: PPC board replacement, please refer to PPC FRU replacement procedure in service manual. - if it is not labelled 2351200-11 rev 001, the board installed on FRU is good: no modification is needed. Shawn Fox 3-24-05 Part #: 2281950 -4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBRPart #: 2281950-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: Added per ECR 2018129, THIS PART IS BACKWARD COMPATIBLE REPLACEMENT FOR 2281950-3, -2, & NO DASH. MS13452 050919 Repairable: No Returnable: No Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: Added per ECR 2018129, THIS PART IS BACKWARD COMPATIBLE REPLACEMENT FOR 2281950-3, -2, & NO DASH. MS13452 050919
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.