GE Healthcare · CTDiscoveryVCT260133224
"Scanner Hardware Stopped Scan" with CAM or Filter errors
Symptom
"Scanner Hardware Stopped Scan" with CAM or Filter errors
Error message
Error: 260133224 The collimator has detected motor slip during z-axis tracking. See long description Mon Nov 9 10:12:31 2009 Error: 260133224 Host: orp Process: orp subsys File: aperture z axis.cxx Line: 169 Mon Nov 9 10:12:30 2009 Host : Collimator Ermes # : 260133224 Exception Class : Secondary Severity : Pri/Soft File : aperture z axis.cxx Line# : 169 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The collimator has detected motor slip during z-axis tracking. WARNING The cams will be re-referenced before the next scan.This may extend the ISD.Cam A Delta : -4323 encoder countsCam B Delta : -1 encoder counts
Applies to
- Models
- Discovery, VCT
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5271284 VCT Collimator Motor and Coupling Assemb See long description for details Part #: 5271284 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: VCT Collimator Motor and Coupling Assembly FRU: Yes Status: Orderable Comments: Per ECR 2052055, 212008898, 4/7/08.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required