GE Healthcare · CTLightSpeedVCT260116086260101018
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Gantry shutdown itself for 1 minute and then reset on its own. GANTRY SHUTDOWN ITSELF FOR 1 MINUTE AND THEN RESET ON ITS OWN BETWEEN SCOUTS AND AXIAL SCANS.
Error message
Error: 260116086 Scanning Hardware Reset is needed to recover See Long description for complete error Tue Oct 31 14:19:54 2006 Error: 260116086 Host: tgp Process: tgp_subsys File: sys_watchdog_err_handle.c Line: 84 Tue Oct 31 14:20:07 2006 Host : Table Gantry Processor Ermes # : 260116086 Exception Class : Safe State Severity : Primary File : sys_watchdog_err_handle.c Line# : 84 Function : No System Function Reported Scan Type : None/Unknown Scan: 9328/2/0 Application ID 211 has failed to report to the system watchdog appId refresh time = 3000 wdog tmo time = 100 Scanning Hardware Reset is needed to recover Error: 260101018 The DCB sub-system failed to return a response to an ORP Tue Nov 22 09:01:31 2011 Error: 260101018 Host: orp Process: orp_subsys File: rcib_msg_reply.c Line: 590 Tue Nov 22 09:01:05 2011 Host : On-Board Rotating Processor Ermes # : 260101018 Exception Class : Abort Severity : Primary File : rcib_msg_reply.c Line# : 590 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The DCB sub-system failed to return a respond to an ORP query message in the allotted time. Expected message code: 508 Timeout: 5000 mS.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required