GE Healthcare · CTLightSpeed24528350003245183179548800053027

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan With A Pink Pop Up Message " Data Acquisition is not ready for scanning.You might have to reboot the system to recover." See Long Description for Details Scanner Hardware Stopped Scan With A Pink Pop Up Message Is Reported Customer cannot perform scanning, the system report scanner hardware stop scan ,and a pink pop up message is reported that:" Data Acquisition is not ready for scanning.You might have to reboot the system to recover."

Error message

Error: 245283 Timeout (6000 ms) waiting on scan disk rezero response from dataacq See long description for full error Mon Nov 26 07:29:05 2007 Error: 245283 Host: svmcct16 Process: scanmgr File: scanmgr events.cxx Line: 1151 Function: System Control : Scan Control Scan Type: Not scanning Scan: 50003/2/0 Scan Seq Id: 0 Timeout (6000 ms) waiting on scan disk rezero response from dataacq.

Error: 245183 Did not receive reply from a request. See long description for a sample error message. Tue Jan 5 09:18:44 2010 Error: 245183 Host: cbct001 Process: scanmgr File: firmware if.cxx Line: 1554 Function: System Control : Scan Control Scan Type: Helical Scan: 17954/3/2 Scan Seq Id: 88000 Did not receive reply from a request. Reply Event : 53027 Reply Source: scan control Error: Timed out while attempting to receive a message.

Error: 200501547 A Communication Error has occurred in uifIpcSendRecvScanRequest See LD for more details Thu Nov 9 08:39:17 2006 Error: 200501547 Host: ct-vct Process: scanRx File: uifIpcMgrScanRx.c Line: 5632 Function: Data Acquisition : OC Processing A Communication Error has occurred in uifIpcSendRecvScanRequest with seq_num = 37 and status = 200000008

Error: 24506

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.