GE Healthcare · Nuclear MedicineMillennium
After upgrade to TTH X-ray tube a VG Hawkeye stopping intermittently during Hawkeye scans with “Stopped on Unknown” on t…
After upgrade to TTH X-ray tube a VG Hawkeye stopping intermittently during Hawkeye scans with “Stopped on Unknown” on the Monitor.
Symptom
After upgrade to TTH X-ray tube a VG Hawkeye stopping intermittently during Hawkeye scans with “Stopped on Unknown” on the Monitor. No other indications or symptoms pointing to a root cause
Error message
ERROR: in AisCheckHVPSerror: AIS WATCHDOG FAULT 8 Tue, Aug 16, 2011 16:56:40 *** ERROR: in AisCheckHVPSerror: AIS WATCHDOG FAULT 8 Exiting from AisCheckHVPSerror... Exiting from AisGetHVPSstate... Exiting from QaqCheckHVPS... Exiting from QaqControlStepCT... Exiting from QaqControlThread... Tue, Aug 16, 2011 16:56:50 *** ERROR: in QdsWindowProc: Acquisition Stopped on unknown !! Exiting from ArmAllocElement... Tue, Aug 16, 2011 16:56:53 * NOTE: in QmnStopMsgErr: ** XRAY acquisition terminated. S 45. Stopped on UNKNOWN (0) Error message "Stopped on Unknown"
Applies to
- Models
- Millennium
- Product
- VGH VG and VG Hawkey and Varicam
- Component
- 2378203 ECL (CDL) Slip Ring Reciever E9300020023 2378203 ECL (CDL) Slip Ring Reciever E9300020023 *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- How do I capture and print COR and Camera Uniformity (CUA) result screens on Genieacq, then send to a processing station for ACR
- Collision and E-stop lights are red on the back of the gantry and no movement with Hand Held Controller (HHC) or the genieacq.
- Need part numbers for the Millennium MG/MC table E-Stop
- How do I Delete a file on Genieacq that the system thinks is still acquiring
- VG, the contour works... but in the opposite way.
- MG_MC_MyoSight_MPR_MPS Bar Phantom Resolution Size