Pular para o conteúdo principal
GE Healthcare · MRSigna2219712221749822186302218501

The Download Of A Prescribed Series Failed Due to Corrupt Protocol.

Symptom

Unable to Scan With The Download of a Prescribed Series Failed Message

Error message

2219712 The executable download of psd failed. Timed out waiting for download complete packet. Try redownloading the CVs

2217498 Timed out waiting for download complete packet. Try redownloading the CVs. 2217498 Host: lx-mr Process: scn File: PsdError.m Line: 324 Timed out waiting for download complete packet. Try redownloading the CVs. Error: 2217498 Host: lx-mr Process: scn File: PsdError.m Line: 324 Timed out waiting for download complete packet. Try redownloading the CVs.

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution