Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT200107023200107002

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner shutdown on its own to the system halted screen. Scanner shutdown on its own to the system halted screen

Error message

Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToScanRequest function. Nov 22 08:55:38 2006 Error: 200107023 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 148 Function: Data Acquisition : OC Processing srx_sm_event_notify: A Failure had been detected while calling srx_sm_generate_action function. Tue Aug 14 16:28:25 2007 Error: 200107023 Host: cccgvct1 Process: scanRx File: srx_sma_manage_scan.c Line: 733 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling SIRx_areScanFilesAvailable function.

Error: 200107002 Invalid event 1014 (SCAN DONE event) for current state 2000 (Idle). See Long Description for complete error. Wed Nov 22 08:55:38 2006 Error: 200107002 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 703 Function: Data Acquisition : OC Processing In

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues