GE Healthcare · CTDiscoveryVCT230000065230000069291443700050003244000
Scanner Hardware Stopped Scan.
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 230000065 Buffer overrun detected during scan data transfer. See long description for full error Error: 230000065 Buffer overrun detected during scan data transfer. This is generally not a hardware issue and will not be remedied by replacing or otherwise servicing the gantry, fiber-optic cable or DIP card. Interrupt bit: 6 Violation count: 1 Scan aborted. Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 3 1 TAXI modem violation 5 1 View length error Mon Dec 15 10:40:11 2008 span style="COLOR: #ff0000"> Error: 230000069 Host: darc Process: dataacq File: hddip.cxx Line: 720 Function: Data Acquisition : Scan Data Acquisition Scan Type: Helical Scan: 29144/9/3 Scan Seq Id: 37000 Two, or more, unexpected events were detected during scan data transfer. Bit Co
Applies to
- Models
- Discovery, VCT
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 10MW25.6 for Discovery CT750 HD 10MW25.6 for Discovery CT750 HD FMI 25426 eSERVICE NOTE SN10121709a : Discovery CT750HD with Veo (MBIR) option Service Introduction NEW PSI CODES CSPH03 HD WITHOUT 128 axial overlap and is MBIR (Veo) compatible CSPH04 HD WITH 128 axial overlap and is MBIR (Veo) compatible ======================================================= Software Collector P/N 5396458 Class C 5397903 - REV 3 Version 10MW26.5 - Discovery CT750 HD Class C - 2010 Release Class M 5397904 - REV 3 Version 10MW26.5 - Discovery CT750 HD Class M - 2010 Release An On-site / Remote identification aid . (Example: check_config) Result from check_config or swhwinfo: 10MW25.6.V40_PS_HD64_G_GTM 10MW25.6.V40_PS_HD64_G_GTL
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.