GE Healthcare · CTLightSpeed183262
LightSpeed 4.X - GOC2
Symptom
"Scan hardware Stopped Scan" with PDU errors "Scan hardware Stopped Scan" with PDU errorss
Error message
Error: 183262 Backup contactor opened without firmware command See the long description for the full error Mon May 17 12:20:18 1999 Error: 183262 Suite: CT01 Host: OBC Process: Generator File: ctvrc_manager.c 1.8 Method: No Method Line:823 Function: XRAY GENERATION : 550 VDC Backup Contactor Intlk Scan: 7073/3/15 Type: Axial Exception Level: Pri/Most Severe Time: 12:20:18:495 Log Series:513 Backup contactor opened without firmware command.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2139289-2 CPDU Control Bd. Replaces part 2139289 Used In CPDU - 2269902. 2139289-2 CPDU Control Bd. for Compact PDU Part #: 2139289-2 Part Type: GP REPAIRABLES Part Weight: 5.35 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 2139289-2 : PDU CONTROL BOARD W/ REW FRU: Yes Status: Orderable Comments: FLS - PDU CONTROL BOARD REWORK ASSEMBLY. THIS P/N 2139289-2, IS P/N T3202KF (2139289) WITH A REWORKED SUB-ASSEMBLY. ADDED PER PCN 220525 BY MS20544 ON 7/17/02. 9/10/02- 2139289-2 REPLACES 2139289 PER PCN 220525. MS12464 11/08/04 SSO# 212001261 FE's are to return defectives and unused/consigned back via FED Ex over night Item #16 A6 panel is the control board
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required