Gradient error 2267894 due to failed XGD Power Supply 5148804-2, also known as XPS.
Symptom
Will Not Scan with Gradient Errors.
Error message
2224666 Auto Prescan failed, Gradients are not ready See Long Description 2224666 Host: lx-bay12b Proc: psc Error: 2224666File: ../PscMainControl.cc Line: 1409Exception Class: Unknown Function: Prescan: Host Autoprescan Auto Prescan failed, Gradients are not ready. 2267894 The X axis XPS-SB is reporting an input bus over-current see long description Tue May 12 16:35:45 2009 Error: 2267894 Host: WBH_RAD_MR11 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The X axis XPS-SB is reporting an input bus over-current.-> Extended Error: Check that 420V input voltage measures correctly. If the fault persists after TPS Resets, and the 420V input power measures correctly, the XPS should be replaced
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Multicoil Bias Fault errors on multiple coils.
- Auto Prescan Failures and RF amp setup incomplete Errors Due to Defective RF amp.
- The system is unable to scan with any surface coil, get dynamic disable faults.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.