GE Healthcare · MRSigna2252306
Scan fails with body TR fault Due to Body Hybrid Splitter.
Symptom
Unable to Scan With T/R driver Fault, Body Path Short
Error message
2252306 TR Driver Fault Detected Short Circuit Source of Fault is Body TR 2252306 TR Driver Fault Detected - Short Circuit. Source of Fault is Body TR. Value: 123 mA Limit: 80 mA->
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2370221 3T DUAL BODY HYBRID aka 3T DUAL BODY T/R SWITCH ASM See Long Description In Renewal Parts, the part name is "3T Dual Body Hybrid". In FEMC the same part number has the part name "3T DUAL BODY T/R SWITCH ASM". Part # 2370221. This part is for the LCC300 short bore 3T both on G3 systems and Excite HD/HDx/HDxT. NOT compatible with the 3T94 long bore 3T.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.