Pular para o conteúdo principal
GE Healthcare · MRSigna222466622247132224725222458022245894062796

The signal is too small due to damaged quad knee coil cables.

Symptom

Auto Prescan Fails With Knee Coil

Error message

2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.

2224713 No successful FTG check in X iterations APS1 will run automatically. Error: 2224713 Host: DMHMRI Process: NSP File: APS:aps_exec Line: 1 No successful FTG check in 2 iterations. APS1 will run automatically.

2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan).

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2383613-3 Medical Advances Quad Knee Coil Cable Assembly MED ADV QUAD KNEE CABLE.
Area
MR