Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18261418260751200

LightSpeed 4.X - GRE

Symptom

Gantry tilt readout no longer matches actual tilt postion of gantry Gantry is showing Parked at 0 Degrees but is actually approximately 8 degrees.

Error message

Error: 182614 Tilt drive is not stopping within 1500 msec. Expected stopping position XX (.01 deg) Current position YY (.01 deg). See Long Description for error details and example Host: ETC Process: Tilt File: tmcShutDown.c 1.31 1/29/93 Line: 121 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 5478/4/0 Type: None/Unknown Exception Level: Pri/Soft Time: 16:25:56:115 Log Series: 151 HW/SW error: Tilt drive is NOT stopping within 1500 msec. Expected Stopping Position -50 (.01 deg). Current Position -23 (.01 deg). Tilt Direction: 0 (0 = BCK 1 = FWD). Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/95 Line: 139 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 7839/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:12:01:160 Log Series: 617 HW/SW error: Velocity tracking error. Actual Velocity 0 (.01 deg/sec) Expected Velocity 100 (.01 deg/sec) Tol 80 (.01 deg/sec) Cur Pos -2181 (.01 deg) Tilt Dir: 1 (0 = BCK 1 = FWD). INFO: Probable Problem could be 180V Power is LOW.

Error: 182607 HW/SW error: Tilt Velocity tracking See LD for more details Thu May 11 23:04:54 2006 Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/9

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X see LD for details P/N 2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X picture below Part #: 2348298 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TIMING BELT, 2.03 PITCH X 6 WIDE FRU: Yes Status: Orderable Parts Offered as: Current: 2348298 Harvested: Not Available Exchange: Not Available Comments: FLS - TIMING BELT. ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19. MS20544 ON 3/24/03
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.