Multi Coil TR Driver Fault.

Symptom

Fails to Scan with TORSO Coil

Error message

2225551 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 1. See Long Description Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225551 File: PMH.m 1.29 Method: spi_tr_mon Line: 6230 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 1. Check cable to multi-coil board or check for an open pin diode.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2215261 1.5T Torso Phased Array Flex Coil See long description for Coil Shipping list and part numbers. This information is Property of GE Healthcare - For GE Healthcare Service Personnel Only. Disclosure to Third Parties Prohibited. From the SIGNA® TORSO PHASED ARRAY COIL REV 2 DIRECTION 2218805 Documentation. To view the most recent appropriate and detailed documentation, click to go to SERVICE METHODS . 1.5T TORSO PHASED ARRAY COIL SHIPPING LIST