Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2601080005005226010102250060

LightSpeed 5.X (60kW)

Symptom

Hardware reset fails with ORP to (CCB, DCB or Jedi)communication errors. Hardware reset fails with ORP to CCB communication errors.

Error message

Error: 260108000 Collimator firmware object: Lost communications with controller GCAN error sending message = 5. See Long Description. Thu Dec 22 16:15:57 2005 Error: 260108000 Host: orp Process: orp_subsys File: ss_comm.c Line: 497 Thu Dec 22 16:15:55 2005 Host : On-Board Rotating Processor Ermes # : 260108000 Exception Class : Abort Severity : Primary File : ss_comm.c Line# : 497 Function : SYSTEM COMMUNICATIONS : Rotating Controller I/F Bus Scan Type : None/Unknown Scan: 50052/1/0 Collimator firmware object: Lost communications with controller GCAN error sending message = 5.

Error: 260101022 Communications between the CCB sub-system and ORP have been lost. See LD for more details Mon Apr 10 07:20:42 2006 Error: 260101022 Host: orp Process: orp subsys File: rcib init state.c Line: 267 Mon Apr 10 07:20:41 2006 Host : On-Board Rotating Processor Ermes # : 260101022 Exception Class : Abort Severity : Primary File : rcib init state.c Line# : 267 Function : No System Function Reported Scan Type : None/Unknown Scan: 50060/6/0 Communications between the CCB sub-system and ORP have been lost. Attempting to establish communications.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2389885-3 (replaces 2389885-2 and 2290254-3) ORP2 BOARD ASSY Positioning HP60 See Long Description for details Part #: 2389885-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ORP2 BOARD ASSY Positioning HP60 FRU: Not Available Status: Orderable Comments: Per ECR 2036188, 212008898, 12/6/2006. Source Pole/Planner Code revised per to supplier confirmation by sso#305010201 2389885-2 ORP2 - Used on LightSpeed 5.X and BrightSpeed (Replaces 2290254-3 per PQR 13030308 ) Part #: 2389885-2 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: ORP2 BOARD ASSY Positioning HP60 FRU: Yes Status: Orderable Comments: Not Available Part #: 2290254-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: P2033WC ORP BD ASSY FRU: Yes Status: Orderable FLS - P2033WC ORP BD ASSY USED ON H-POWER, ORP ASSY
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.