Pular para o conteúdo principal
GE Healthcare · MRSigna222385010303

Autoprescan Intermittently Never Ends With Errors 2249352, 2223850, and 2249425 Logged due to STIF Board.

Symptom

Auto prescan Intermittently Never Ends With Errors 2249352, 2223850, and 2249425 Logged See Long Description Intermittently, auto prescan never ends, the gradients do not pulse, nothing happens, and the errors shown below are found logged in the /usr/g/service/log/gesys* message log. This happens about once per day. Thu Aug 6 08:16:42 2009 Error: 2249352 Host: lx-mr Process: diagCoord File: cm_rcalls.c Line: 1475 Task ID: 0x6735280 Time: 279982 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Maximum CM connect failure with these arguments: (00000045, ee 02 2707 0a000102, 00000010) Thu Aug 6 08:16:42 2009 Error: 2223850 Host: lx-mr Process: NSP File: PMH.cc Line: 490 Exception Class: Unknown Function: Host: NSP Contents of packet can''t be routed to a host application: opcode:835 seq:289 rev:2 type:1 length:62 stat:-1 resp:ASHIM_VRE req:ASHIM_APS First 16 bytes usr data:000000000000000000000001000000040000000009b90000ffff44a8ffff44a8. Thu Aug 6 08:16:44 2009 Error: 2249425 Host: lx-mr Process: diagCoord File: excLog.c Line: 1777 Task ID: 0x6eca750 Time: 280174 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A A network error was detected by task tCM_TxD3 (0x6735280). Additional information in /usr/g/service/log/vxNetLg-44623->

Error message

2223850 Content Of Packet Cant Be Routed To A Host Application Thu Feb 2 11:01:58 2006 Error: 2223850 Host: RIMIMR Process: NSP File: PMH.cc Line: 456 Contents of packet can't be routed to a host application: opcode:10303 seq:20 rev:2 type:1 length:8 stat:-1 resp:WavRec req:HC

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution