Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260116313260101006260100137

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 260116313 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out. see LD Tue May 16 22:03:29 2006 Error: 260116313 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 349 Running: System Reset Tue May 16 21:37:53 2006 Host : Table Gantry Processor Ermes # : 260116313 Exception Class : Safe State Severity : Secondary File : subsys_mgr.c Line# : 349 Function : SYSTEM COMMUNICATIONS : Control Network Scan Type : None/Unknown Scan: 0/0/0 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out.

Error: 260101006 The DCB has failed to respond to an exception in the allotted time Mon Oct 15 9:36:44 2007 Error: 260101006 Host: orp Process: orp_subsys File: rcib_except_state.c Line: 269 Fri Sep 8 07:32:29 2006 Host : On-Board Rotating Processor Ermes # : 260101006 Exception Class : Soft Severity : Pri/Soft File : rcib_except_state.c Line# : 269 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The DCB has failed to respond to an exception in the allotted time. A reset of the controller may be required. Notification group number: 1 Notification group level: 8

Error: 260100137 Host caused the firmware subsystems to SHUTDOWN

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
2404320-2 CONSOLE INTERCOM PRESCRIBED-TILT HUB See Long Description for More ... 2404320-2 CONSOLE INTERCOM PRESCRIBED-TILT HUB Part #: 2404320-2 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: CONSOLE INTERCOM PRESCRIBED-TILT HUB FRU: Yes Status: Orderable Comments: Per ECR 2062448, 212008898, 10/8/08. ** FE ASSIST ** CT/PET Discovery System's must use the P/N 2404320. CT System's must have FMI 25407 or FMI22918 for RT16/Xtra performed prior to installing 2404320-2. Verify Software Configuration before ordering/installing this FRU. To do this, open a Unix Shell and verify one of the following base software versions is installed: ctuser@hostname} swhwinfo -v 07MW11.10_SP2-0-26.Hardware info / 07MW18.4_SP2-0-26.Hardware info / 07BW08.4_SP2-0-26.Hardware info / 08BW17.7 Hardware info / 08BW20.x Hardware info (212001261 10/21/08)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required