GE Healthcare · MRSigna2241510
GPM Power Supply Fault Due to GPM Power Supply.
Symptom
Will Not Scan with Gradient Errors.
Error message
2241510 The GIP is reporting GPM Power Supply Power Off. Mon Jun 25 23:08:14 2001 Error: 2241510 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 1882 The GIP is reporting GPM Power Supply Power Off. Check the GPM Power Supply circuit breakers and incoming power. Check the cable between the GIP and the GPM Power Supply.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2154416-4 GPM POWER SUPPLY WITH FRU PACKAGING 2154416-4 GPM POWER SUPPLY WITH FRU PACKAGING
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.