GE Healthcare · CTLightSpeed260134703903212897741751380
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
The table fails to move up or down but the elevation readout numbers on the gantry display do change whenever a move button is pressed and somethimes the readout shows a negative value.
Error message
Error: 260134703 Elevation moved without command. See LD for more details Wed Jun 14 12:48:34 2006 Error: 260134703 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 530 Wed Jun 14 12:48:34 2006 Host : Table Sub-System Ermes # : 260134703 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 530 Function : No System Function Reported Scan Type : None/Unknown Scan: 412 Elevation moved without command. If characterize completed appears above, this isnt error. If this appears many times, check potentiometer. Current position = 903212 um Last stop position = 897741 um Tolerance = 5000 um Time since last stop = 751380 msec
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5127774 ELEVATION POTENTIOMETER ASSY-VCT See Long Description for details Part #: 5127774 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION POTENTIOMETER ASSY FRU: Yes Status: Orderable Comments: source pole/planner code revised per to Global Table transfer
- Area
- CT
Full cause & step-by-step resolution