GE Healthcare · CTLightSpeedVCT260134703903212897741751380260125041848952
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Intermittent Scan Hardware Stopped Scan message WITH Error: 260003158 The Filter Home Flag Has Moved Since The Last Reference, and, Error: 260003141 Filter homing problem.
Error message
Error: 260134703 Elevation moved without command. See LD for more details Wed Jun 14 12:48:34 2006 Error: 260134703 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 530 Wed Jun 14 12:48:34 2006 Host : Table Sub-System Ermes # : 260134703 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 530 Function : No System Function Reported Scan Type : None/Unknown Scan: 412 Elevation moved without command. If characterize completed appears above, this isnt error. If this appears many times, check potentiometer. Current position = 903212 um Last stop position = 897741 um Tolerance = 5000 um Time since last stop = 751380 msec Error: 260125041 Uncommanded Table Motion with the Elevation drive. See long description for a sample error message. Thu Jan 28 08:29:19 2010 Error: 260125041 Host: tgp Process: tgp subsys File: ta callback gt.c Line: 504 Thu Jan 28 08:29:19 2010 Host : Table Gantry Processor Ermes # : 260125041 Exception Class : Abort Severity : Primary File : ta callback gt.c Line# : 504 Function : No System Function Reported Scan Type : Axial Scan: 1212/2/9 Uncommanded Table Motion with the Elevation drive. Current pos:848952 um Last stop
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5306599 (replaces 5132669-2 and 5132669-3) Step Motor Driver Positioning GT P - used on GT and VT tables See Long Description for details Part #: 5306599 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: Per Kittyhawk RSPL, 212008898, 6/9/08 Part #: 5132669-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Part #: 5132669-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: added per to PET Table RSPL. source pole/planner code revised per to Global Table transfer
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.