Pular para o conteúdo principal
GE Healthcare · MRSigna2248259221863022185012248163

The Download of a Prescribed Series Failed Due to Remote RF Master Exciter.

Symptom

The Download Of A Prescribed Series Failed

Error message

2248259 VERY LOW SIGNAL STRENGTH IN RRF CAL. Low or no Signal Strength Seen From Exciter 0 (0=Master,1=Slave) Mon May 3 13:52:03 2004 Error: 2248259 Host: lx-mr Process: NSP File: APS:RrfCalibrator::excCal Line: 2541 VERY LOW SIGNAL STRENGTH IN RRF CAL Low or no signal strength seen from Exciter 0 (0=Master,1=Slave) during RRF cal.

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

2248163 Receiver calibration validation error. 2004 Error: 2248163 Host: OR3T Process: NSP File: APS:RrfCalibrator::rcvCal Line: 4036 Receiver calibration validation error. Significant measurement discrepancies exist between receivers. First detected: Rx1 vs. Rx2

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues