Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2100004661308589775512866210000447165801308567210

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Images are showing up on the monitor with autoview but not in Image Works. Archiving images and networking of the images are failing.

Error message

Error: 210000466 Info: stat() was successful for file /usr/g/sdc image pool/reserved/IMSsefB1UTz/E* See long description for a sample error. Mon Jun 20 13:09:35 2011 Error: 210000466 Host: cter Process: dms File: DMFInstall.cxx Line: 417 Info: stat() was successful for file /usr/g/sdc image pool/reserved/IMSsefB1UTz/E6717U0 S2U0 I1 U0.1308589775 512866

Error: 210000447 Error: unable to connect to session IOSTERRA xxxxxx with error code -2 error message Session Not Connected Mon Jun 20 13:09:37 2011 Error: 210000447 Host: cter Process: dms File: DMFInstall.cxx Line: 229 Error: unable to connect to session IOSTERRA 16580 1308567210 51895 with error code -2 error message Session Not Connected

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues