Symptom

Scanning aborted midway through exam with scan communication errors.

Error message

Error: 182126 LAN Watchdog TIMEOUT with OBC (uuid 193) See LD for complete error Wed Sep 20 08:39:07 2006 Error: 182126 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 851 Function: No System Function Reported Scan: 50091/8/0 Type: None/Unknown Exception Level: Secondary Time: 08:39:01:815 Log Series: 724 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. (Last Wdog Probe Sent:491163 Last Response Rcvd:490966 (in Ticks)).

Error: 182116 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). See Long Description for Error Mon Dec 6 22:19:40 2004 Error: 182116 Suite: CT01 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Method: No Method Line: 766 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:19:33:885 Log Series: 27 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. Will need a Hardware Reset. Or Tue Sep 23 19:48:54 2003 Error: 182116 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 795 Function: No System Function 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.