"Scan hardware Stopped Scan" with collimator errors.

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260133169 drive monitoring detected the current motor position is out of the tolerance band for a successfull move. See long description for complete error Mon Jan 29 13:54:09 2007 Error: 260133169 Host: orp Process: orp_subsys File: ccb_drive_timers.cxx Line: 599 Mon Jan 29 13:54:08 2007 Host : Collimator Ermes # : 260133169 Exception Class : Abort Severity : Secondary File : ccb_drive_timers.cxx Line# : 599 Function : No System Function Reported Scan Type : Scout Scan: 7295/1/1 CAM B drive monitoring detected the current motor position is out of the tolerance band for a successfull move Start Position = 5944um 0cnt Desired Position = 1219um 1307cnt Current Motor Position = 2105um 1064cnt Current Encoder Position= 4870um 323cnt Retry Tolerance = 150um Tue Dec 9 16:05:36 2008 Error: 260133169 Host: orp Process: orp_subsys File: ccb_drive_timers.cxx Line: 599 Tue Dec 9 16:05:35 2008 Host : Collimator Ermes # : 260133169 Exception Class : Abort Severity : Secondary File : ccb_drive_timers.cxx Line# : 599 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 . Filter A drive monitoring detected the current motor position is out of the tolerance band for a 

Applies to

Models
Discovery, LightSpeed
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5335448 VCTHD Failsafe CCB Asm Part #: 5335448 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCTHD Failsafe CCB Asm FRU: Yes Status: Orderable FE can use existing mounting hardware on the system
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.