GE Healthcare · Nuclear MedicineMillennium
VG/VH reported ERROR #46 on gantry display
Symptom
VG/VH reported ERROR #46 on gantry display
Error message
Error #46 (Error 46) GANTRY_ERROR_MCH_PORT_168 - MCH safe mode mismatch (Port 168) Error #46 (Error 46) GANTRY_ERROR_MCH_PORT_168 MCH safe mode mismatch (Port 168)
Applies to
- Models
- Millennium
- Product
- VGH VG and VG Hawkey and Varicam
- Component
- *Gantry Gantry Gantry 5197979 (PCA000492) MCH Board for SPARE PART 5197979 (PCA000492) MCH Board for SPARE PART 5197979 MCH Board is not compatable with VG motion software version 6.0AH Verify VG Gantry Firmware version. VG Gantry Firmware Version must be 6.0P7 . See KOID 153674
- Area
- Nuclear Medicine