Pular para o conteúdo principal
GE Healthcare · Nuclear MedicineMillennium

VG/VH reported ERROR #46 on gantry display

Symptom

VG/VH reported ERROR #46 on gantry display

Error message

Error #46 (Error 46) GANTRY_ERROR_MCH_PORT_168 - MCH safe mode mismatch (Port 168) Error #46 (Error 46) GANTRY_ERROR_MCH_PORT_168 MCH safe mode mismatch (Port 168)

Applies to

Models
Millennium
Product
VGH VG and VG Hawkey and Varicam
Component
*Gantry Gantry Gantry 5197979 (PCA000492) MCH Board for SPARE PART 5197979 (PCA000492) MCH Board for SPARE PART 5197979 MCH Board is not compatable with VG motion software version 6.0AH Verify VG Gantry Firmware version. VG Gantry Firmware Version must be 6.0P7 . See KOID 153674
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare Nuclear Medicine issues