GE Healthcare · MRSigna2244655

Will Not Scan with Gradient Errors Due to the GSC Twin Gradient Switch Module Box When Switching Modes

Symptom

Scanner Pausing Between Series when Switching from Whole to Zoom Mode

Error message

2244655 Gradient Switch Fault Detected. Scan will be paused 2244655 Host: TwinExciteLXMR Process: NSP File: SCP:Grad Line: 704 Gradient Switch Fault Detected. Scan will be paused.-> Extended Error: Please restart the scan (continuing the scan might cause image quality issues). If the fault persists,replace the Gradient Switch

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
Area
MR