Pular para o conteúdo principal
GE Healthcare · MRSigna225478213444

The UPM Calibration files are corrupt and create an unrecoverable condition.

Symptom

Unable to scan with Power Monitor Faults

Error message

2254782 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Mon Aug 14 18:50:31 2006 Error: 2254782 Host: MMCLXMR1 Process: NSP File: SCP:UPM Line: 657 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Peak value detected = 13444 W.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues