GE Healthcare · CTLightSpeedVCT26011801150075260118326260118315
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118011 CAN_WRITE_OVERRUN. This error occurs when the library waits more than CAN_DATA_MUTEX_TIMEOUT to be able to send a message See LD for complete error Fri Feb 3 07:59:35 2006 Error: 260118011 Host: orp Process: orp_subsys File: gen_exec.c Line: 459 Fri Feb 3 07:59:34 2006 Host : On-Board Rotating Processor Ermes # : 260118011 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 459 Function : SYSTEM COMMUNICATIONS : Generator CAN Serial Link Scan Type : None/Unknown Scan: 50075/6/0 CAN_WRITE_OVERRUN. This error occurs when the library waits more than CAN_DATA_MUTEX_TIMEOUT to be able to send a message. Error: 260118326 The communication watchdog with the generator has failed. See long description for a sample error. Thu May 27 07:10:44 2010 Error: 260118326 Host: orp Process: orp subsys File: gen exec.cxx Line: 7521 Exception Class : Abort Severity : Primary Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The communication watchdog with the generator has failed. Error: 260118315 JEDI ID Communication Timer Expired. Check power and commnication interconnects. See Long Description for complete error Mon Dec 6 12:04:49 2004 E
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5136161-4 PPC KV Control replaces (5136161, 5136161-2, 5117564-10) used by all 5.X, Brightspeed and VCT gantries See Long Description for Details PPC KV Control Bd. 5136161-4 replaces (5136161, 5136161-2, 5117564-10) this is the orderable Programmed KV Control Board for HP60, LSPro 80KW , 100KW and VCT 7.X, RT16, BrightSpeed Power and Select Systems (Velocity, Delight, Meri) Part #: 5136161-4 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL P6.33BF6 AS FRU FRU: Yes Status: Orderable Current: 5136161-4 Harvested: Not Available Exchange: Not Available Comments: Jedi KV Control Board used on Definium 8000, Revolution XR/d-2X (Lightning), Definium 6000, and Precision 500D. Also may be used on all Innova systems Bolus/Unity/SPAD/Harmony. P.Liebau 1/2010 Part #: 5136161-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED kV CONTROL 6.30BFA AS FRU FRU: Yes Status: Orderable Comments: In order to avoid Jedi software download issues on Definium 8000, Revolution XR/d-2X (Lighting), Definium 6000, and Precision 500D you may order 5136161-4, KV Control Board developed for these systems. Part #: 5136161 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED KV CTRL P6.30BFA as FRU FRU: Yes Status: Orderable Comments: (2/21/06 212001261) The 5136161 KV board replacement is the preferred part used on all HP60 hp 80 LS-RT and VCT products.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required