Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT230017145230020003230017165230017078503421111522307

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

System fails to complete detailed calibrations. System will still perform routine scanning ok. System fails to complete detailed calibrations. System will still perform routine scanning ok.

Error message

Error: 230017145 ACCS: Calculation(CAL_PROC_BIS_IBO) failed with 0 See long description for sample error message. Wed May 11 10:00:55 2011 Error: 230017145 Host: medxct1 Process: accs File: aircal_spectral_bis_ibo.c Line: 206 ACCS: Calculation(CAL_PROC_BIS_IBO) failed with 0

Error: 230020003 Bad channel found at row: 14, channel: 424 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Mon Jul 31 22:23:16 2006 Error: 230020003 Host: ct64_309 Process: accs File: cal_proc_pb.c Line: 437 Bad channel found at row: 14 , channel: 424. Click Here to Decode the ROW NUMBER

Error: 230017165 ACCS: Air Calibration of Spectral calculation(BIS IBO) failed. See long description for a sample error message. Wed May 11 10:00:55 2011 Error: 230017165 Host: medxct1 Process: accs File: aircal_spectral.c Line: 1025 ACCS: Air Calibration of Spectral calculation(BIS IBO) failed.

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microsec

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5134959 FRDM Module L-FINE 32 replaces 5123019 Used on VCT 32 Detector See Long Description for Details. 5134959 FRDM Module L-FINE 32 replaces 5123019-Used on VCT 32 Detector Part #: 5134959 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: L-FINE 32 Detector Module Kit FRU: Yes Status: Orderable Part #: 5123019 Part Type: MST DEFAULT LOAD Part Weight: 0.00 LBR Repairable: Not Available Returnable: Not Available Hazardous Material: N Description: L-FINE 32 DIGITAL MODULE ASSY MP3 FRU: Yes Status: Non Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues