Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT1111710008

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

CU does not want the Dose Structured Report, Series 997, to be generated on their exams. See long description for details CU does not want the Dose Structured Report, Series 997, to be generated on their exams. What is a dose structured report ? It is a structured report object, which complies with DICOM Standard 3.16 for dose reporting. The DICOM fields in the report can be manually exported to a data base to track patient dose.

Error message

Error: 11117 Remote does not support push for SOP_Class_UID See Long Description Two version of this error Exist. “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.67 , so all images with this SOP_class_UID have been rejected Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.22 , so all images with this SOP_class_UID have been rejected “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 * SR stand for "Structured Report"

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Software
09MW08.10 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) June 2009 Inital M3 release. This software is now obsolete, this site should be upgraded to 11HW12.5 software released via FMI 25427 09MW08.10 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) June 2009 Inital M3 release. LightSpeed 7.2 ( 09MW08.10 ) Software Release supports GOC6 consoles used on VCT 64, VCT XT, VCT XTe and VCT Std. - LightSpeed 7.2 (09MW08.10 )Software for GOC6 Collector 5328048 Rev 3 Includes OS and Application CD's - Operating System Linux 5.3.8 Software CD 5323584 Rev 1 - Application LightSpeed VCT GOC6 DVD 5336533 Rev 3 - Class C LightSpeed 7.2 5336535 Rev 2 - Class M LightSpeed 7.2 5336536 Rev 2 An On-site/Remote identification aid . (Example: check_config) Result from check_config or swhwinfo 09MW08.10.V40_H_H64_G_GTM 09MW08.10.V40_H_H64_G_GTM Please create an ITRAK report for any software related issues
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues