LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Pink Pop-up Scan Disk Array has encountered a hard drive failure. If another hard drive fails you will lose scan data. See Long Description: Pink Pop-up message is displaced to the customer on boot up. Scan Disk Array has encountered a hard drive failure. If another hard drive fails you will lose scan data. "Scan Disk Array has encountered a hard drive failure. The System is functional, but if another hard drive fails you will lose scan data. Please Contact GE Service to have the disk array repaired as soon as possible."
Error message
Error: 230000900 WARNING slow write to disk See llong decsriptions for full error Thu Jan 29 08:54:34 2009 Error: 230000900 Host: darc Process: dataacq File: file_save_thread.cxx Line: 194 WARNING slow write to disk: 51.538526 sec to write 33554432 bytes; scan CT07/152/3/3-1233218126/727031; scan seq 75000 Example: This took 51 seconds to write 33 mb. which is extremely long. 51.538526 sec to write 33554432 bytes; Error: 245275 WARNING- Scan setup took 4479 ms overall See long description Thu Oct 29 07:34:53 2009 Error: 245275 Host: vct Process: scanmgr File: scanmgr_lib.cxx Line: 1283 Function: System Control : Scan Control Scan Type: Cine Scan: 0/0/0 Scan Seq Id: 1212000 WARNING- Scan setup took 4479 ms overall. This could potentially cause scan timing issue or scan abort. Error: 230011074 Unable to open a scan file on the scan data disks. Read of scan file header failed in lseek() call See LD for error Mon Nov 8 11:14:52 2004 Error: 230011074 hmaich_gels_ct4 scanmgr sdio_xfs_io_ops.c 625 Function: Data Acquisition : SBC Processing Scan Type: Helical Scan: 6826/2/1 Scan Seq Id: 4000 Unable to open a scan file on the scan data disks. Read of scan file header failed in lseek() ca
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible - used On GOC6 Disk Array See Long description: Part #: 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible used on GOC6 Disk array ( 5308759 ). Part #: 5308760 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible with 5308759 FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 7/15/08. ** FE ASSIST ** This is a single Hard Drive used in the High Speed Disk Array on the GOC6 Console CT 750 HD and VCT (212001261 23-Jan-2009) ST32050310NS = 250 GB SATA II Hard Disk Drive
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.