Symptom

Gantry will not rotate

Error message

Error: 183944 Axial Encoder Feedback Error Encoder counts dropped during scan, may result in rotated images. Check images for proper orientation. Call Service. See Long Description for complete error Wed Nov 16 10:18:38 2005 Error: 183944 Suite: CT27 Host: STC Process: Axial File: ax_error.c Method: No Method Line: 383 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 41674/201/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:18:39:495 Log Series: 764 Axial Encoder Feedback Error Encoder counts dropped during scan, may result in rotated images. Check images for proper orientation. Call Service.

Error: 183908 Axial velocity ACCEL tracking fault See the Long Description for the full error code Mon Aug 27 09:12:45 2007 Error: 183908 Host: STC Process: Axial File: ax_error.c Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 14756/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 09:12:45:860 Log Series: 363 Axial velocity ACCEL tracking fault. current: commanded = 39990 mA current: D/A measured = 39844 mA PDU measured = 39453 mA (HW/SW Error: Retry Rx to verify probable Axial Driver or. Axial Encoder hardware problem.)

Error: 183921 Axial condit

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-198495P1 Axial Belt PITCH 1/2 WIDTH 02.000 LG 170.000 See Long Discription For Details Part #: 46-198495P1 Part Type: GP CONSUMABLES Part Weight: 2.27 LBR Repairable: No Returnable: No Hazardous Material: N Description: PITCH 1/2 WIDTH 02.000 LG 170.000 FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.