Symptom

Artifact on head images. See long description for artifact. Artifact on head images

Error message

Error: 290012001 There are more than XX consecutive bad views - no correction applied . . . exam/series/scan = XXX / XX / XXXX See LD for complete error Thu Feb 9 08:53:27 2006 Error: 290012001 Host: ig3 Process: image_generation File: MissedView.cpp Line: 910 View correction for view 9300 from views 9278, 9322 There are more than 21 consecutive bad views - no correction applied exam/series/scan = 345/4/1

Error: 290012000 View corrections were applied in view # The errror 290012000 "View corrections were applied in view number" message indicate the Image Generator or recon software detected a corrupt view and has applied view corrections to image. This Error alone does not indicate a problem. See service note SN05102815a. Wed Feb 13 13:46:25 2008 Error: 290012000 Host: ig3 Process: image_generation File: MissedView.cpp Line: 129 View corrections were applied in view 5400. exam/series/scan = 345/2/1 From SN05102815a Error code 290012000 is a status notification that Recon has made a correction to maintain image quality. This error is transparent to the Customer and has no effect on system performance. This condition is generated when the IG reads in the scan data and checks the KV 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 See Long Description for details 2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 Part #: 2246223-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HV TANK HEMIT JEDI 60DC (SPARE) FRU: Yes Status: Orderable Comments: Not Available