GE Healthcare · MRSigna2258103222362530011
Coil id not recognized intermittently due to Coil Cable
Symptom
Coil ID not recognized intermittently. Only one coil is affected.
Error message
2258103 Cannot send CoilID Pkt: OP SCT CONNECT CODE TYPE to SCP, Connector: 3, Pls do a tps reset & try again Fri May 13 11:21:04 2011 Error: 2258103 Host: SJMR-OC0 Process: coilCK File: coilCK scp.c Line: 1643 Exception Class: Unknown Function: Host: Coil Check Cannot send CoilID Pkt: OP SCT CONNECT CODE TYPE to SCP, Connector: 3, Pls do a tps reset & try again 2223625 Send packet opcode:30011 seqnum:11 rev:2 type:1 length:0 req:SCAN_01_NOR resp:COILCK status:0 failed. 2223625 lx-mr scn mrmailmgr.c 791 Exception Class: Unknown Function: Software/Firmware: Host Communications Send packet opcode:30011 seqnum:11 rev:2 type:1 length:0 req:SCAN_01_NOR resp:COILCK status:0 failed.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2416711 FRU POSTERIOR CABLE, 1.5T CARDIAC 2416711 FRU POSTERIOR CABLE, 1.5T CARDIAC. 2416710 SYSTEM CABLE, 1.5T CARDIAC Part #: 2416710Part Type: GP CONSUMABLES Part Weight: 0.00 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: FRU SYSTEM CABLE, 1.5T CARDIACFRU: YesStatus: Orderable
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.