LightSpeed 5.X (60kW)

Symptom

Gantry control pad pushbuttons are depressed into the pad

Error message

Error: 260111048 Gantry Control button hardline is active at gantry start-up. Mon Aug 9 15:09:39 2010 260111048 2 tgp tgp_subsys guif_can_mgr2.c 6215 Mon Aug 9 15:09:40 2010 Host : Table Gantry Processor Ermes # : 260111048 Exception Class : Secondary Severity : Pri/Soft File : guif_can_mgr2.c Line# : 6215 Function : OPERATOR INPUT/OUTPUT : Gantry Operator Interface Scan Type : None/Unknown Scan: 0/0/0 Gantry Control button hardline is active at gantry start-up. If nobody touched buttons at start-up, any buttons must be stuck It disables fail-safe logic for gantry tilt and table motion. Replace Gantry Control Panel until this message disappears.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution