GE Healthcare · MRSigna2254786
UPM Errors During Scan Due to Defective 3T Dual Body Hybrid
Symptom
UPM Errors During Scan
Error message
2254786 The Universal Power Monitor 1 Has Detected an Invalid Configuration. Source of Fault is NB Detector Board RF Cable Missing. See Long Description for full error and extended log... NOTE: The errors below were copied from a 3.0T 14xM5 system: Date Time ErrorCode Process File Line Message Jan 24 Thu 17:11:21 2254786 NSP SCP:UPM 894 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Unknown Coil: Unknown Scan: -1/-1 The Universal Power Monitor 1 Has Detected an Invalid Configuration. Source of Fault is NB Detector Board RF Cable Missing.-> Jan 24 Thu 17:11:21 2254786 NSP SCP:UPM 894 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Unknown Coil: Unknown Scan: -1/-1 The Universal Power Monitor 2 Has Detected an Invalid Configuration. Source of Fault is NB Detector Board RF Cable Missing.-> Extended Error Log: 2254786 ERROR: EM_UPM_CONFIG_FAIL The %b Has Detected an Invalid Configuration.\b\ Source of Fault is %b. Extended Error: For each of the following failures look for the cables connected in the locations indicated below: UPM1 NB Detector Board RF Cable Missing: J1-J6 in slot 14(ASC),slot 20L(CAM). UPM2 NB Detector Boa
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2370221 3T DUAL BODY HYBRID aka 3T DUAL BODY T/R SWITCH ASM See Long Description In Renewal Parts, the part name is "3T Dual Body Hybrid". In FEMC the same part number has the part name "3T DUAL BODY T/R SWITCH ASM". Part # 2370221. This part is for the LCC300 short bore 3T both on G3 systems and Excite HD/HDx/HDxT. NOT compatible with the 3T94 long bore 3T.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.