TR Driver Fault Due to Head Coil.

Symptom

Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.

Error message

2252306 TR Driver Fault Detected Short Circuit Source of Fault is Head TR See Long Description for Extended Error. 2252306 ERROR: EM_DCB_TR_SHORT_CKT TR Driver Fault Detected - Short Circuit. Source of Fault is %b. Value: %d mA Limit: %d mA Extended Error: If this error happens repeatedly then check for a short-circuited PIN diode or other short-circuit condition somewhere in the failing transmit chain. Use TR DD MC System Path diagnostic or Gather Load Data and verify that the short circuit values are less than 30mA. If error happens intermittently and cannot be traced to failed hardware then it may be due to gradient noise in the system.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2341973 1.5T SPLIT HEAD COIL ASSEMBLY 1.5T Split Head Coils Part Number 2341973 1.5T Split Head Coils Part Number 2341973
Area
MR