The issue is due to an incorrect permission for the device file used to communicate. 15M4B service note on new service p…

The issue is due to an incorrect permission for the device file used to communicate. 15M4B service note on new service procedure for HOS RRI supply.

Symptom

RRI Shim Supply communication failure on 15M4B Software RRI Shim Supply communication failure on 15M4B Software.

Error message

HOS calibration message Opening of RS-232 port failed for shim supply. see long description The following error message is witnessed: Opening of RS-232 port failed for shim supply. Please check the cable connections to the shim supply or disable the communication to Shim supply using Guided Install, if shim supply is not connected.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.