Will Not Scan With RF Errors due to the Erbtec Processor Board.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2225607 ERBTEC RF Amplifier fault 16 HPC PORT ID Error: 2225607 File: PMH.m 1.29 Method: spi rf mon Line: 3035 ERBTEC RF Amplifier fault 16. 48V power supply unacceptable. Less than 42.9V or greater than 51.5V. Possible defective Power Control Board or Solid State Amplifier Board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-305000P4 Erbtec Amplifier PROCESSOR BOARD For 1.0/1.5T Amplifier NOTE 1. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD.