GE Healthcare · MRSigna2225607
Will Not Scan With RF Errors due to the Erbtec Processor Board.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2225607 ERBTEC RF Amplifier fault 16 HPC PORT ID Error: 2225607 File: PMH.m 1.29 Method: spi rf mon Line: 3035 ERBTEC RF Amplifier fault 16. 48V power supply unacceptable. Less than 42.9V or greater than 51.5V. Possible defective Power Control Board or Solid State Amplifier Board.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-305000P4 Erbtec Amplifier PROCESSOR BOARD For 1.0/1.5T Amplifier NOTE 1. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.