GE Healthcare · CTLightSpeed26013470779382279400558796050189418827
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
Intermittent cradle drive error and table reset during cradle or elevation movement Cradle (intermittently) starts with a jerky movement, immediately followed by a table power cycle. After about 15 sec the table is available again for a movement. The table power cycle can also occur (intermittently) during lowering the table if the cradle (not IMS) was not at home position before down movement.
Error message
Error: 260134707 Elevation could not stop with stop command. If table is moving downward, check valve See Long Description for error versions Mon Sep 19 08:02:03 2005 Error: 260134707 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 616 Mon Sep 19 08:02:03 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 616 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 793822 um current position = 794005 um commanded direction = UP time from stop = 587960 ms Mon Dec 12 10:22:24 2005 260134707 Host: tgp tgp_subsys gtcb_emc_movechk.cxx 579 Mon Dec 12 10:22:24 2005 Host : Table Sub-System Ermes # : 260134707 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 579 Function : No System Function Reported Scan Type : None/Unknown Scan: 50189/1/0 Elevation could not stop with stop command. If table is moving downward, check valve. last position = 418827 um current position = 418382 um commanded direction = DOWN time from stop = 1040 ms
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5127284-3 (replaces 5127284) PROPORTIOAL VALVE ASSY Positoning GT - FRU yes - this assembly contains two proportional valves. See Long Description for details and picture: Part #: 5127284-3Part Type: GP CONSUMABLES Part Weight: 0.00 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: PROPORTIOAL VALVE ASSY Positoning GTFRU: YesStatus: Orderable Comments: added per to ECO2079401 by sso#305010201 in Jun 8, 2009 Part #: 5127284 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROPORTIOAL VALVE ASSY Positoning GT FRU: Yes Status: Orderable Comments: Not Available This part is actually an assembly - the assembly contains two proportional valves and wire cable harness assembled/wired together. GT
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required