Symptom

"Scanner Hardware Stopped Scan" with CAM or Filter errors

Error message

Error: 260133156 CAM A/B Move Timeout Timer expired - move did not complete in time See long description for full error Mon Sep 29 07:35:25 2008 Error: 260133156 Host: orp Process: orp_subsys File: ccb_drive_timers.cxx Line: 362 Mon Sep 29 07:35:23 2008 Host : Collimator Ermes # : 260133156 Exception Class : Abort Severity : Primary File : ccb_drive_timers.cxx Line# : 362 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 CAM A Move Timeout Timer expired - move did not complete in time

Error: 260133155 CAM (A or B) Motion With Out Command Exceeded The Error Tolerance See Long Description for error detail Sun Jul 11 21:12:37 2010 Error: 260133155 Host: orp Process: orp_subsys File: ccb_drive_timers.cxx Line: 298 Sun Jul 11 21:12:35 2010 Host : Collimator Ermes # : 260133155 Exception Class : Abort Severity : Secondary File : ccb_drive_timers.cxx Line# : 298 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 . CAM B Motion With Out Command Exceeded The Error Tolerance. Desired Position : 5755um 0cnt Current Position : 5817um -19cnt Error Tolerance: 60um Sun Mar 29 11:16:32 2009 Error: 260133155 Host: orp Process: orp subsys File: ccb

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5271284 VCT Collimator Motor and Coupling Assemb See long description for details Part #: 5271284 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: VCT Collimator Motor and Coupling Assembly FRU: Yes Status: Orderable Comments: Per ECR 2052055, 212008898, 4/7/08.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.