GE Healthcare · MRSigna2262134

Coil Will Not Properly Connect and is not recognized when plugged in Due to Bent Pins in Coil Connector.

Symptom

Coil Will Not Properly Connect and is not recognized when plugged in Coil Will Not Properly Connect and is not recognized when plugged in.

Error message

2262134 Coil validation error. Coil ID device was not detected on a coil. Coil ID is required for this connector See Long Description Tue Sep 2 07:14:13 2008 Error: 2262134 Host: hde Process: NSP File: SCP:SRI Line: 348 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 Coil validation error: Coil ID device was not detected on a coil. Coil ID is required for this connector. Please try connecting coil again. Extended Error: This could be a problem with the coil ID communication chain which includes the Coil, the cable between the Coil and the LPCA, the LPCA, the cable between the LPCA and SRI, and the SRI. Please run Coil ID diagnostics.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution