GE Healthcare · MRSigna2252106

2252106 Failed to determine hardware state for AGP Due to Both MGD power supplies.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung.

2252106 Failed to determine hardware state for APS board within 20 seconds. Error: 2252106 Host: BARLOW_MR Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for APS board within 20 seconds The board may be busy loading vxWorks or may be hung. Also check the serial cable for the board and the serial terminal server.

2252106 Failed to determine hardware state for SCP board within 20 seconds

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5181495-3 400W Quad Output 6Ux8HP cPCI Power Supply This is a replacement for EOPL 2294300-8
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.