Application does not Boot up with VRE Errors due to ICN
Symptom
Can Not Boot Application Software With VRE Errors See Long Description 2249418 Host: LJP MR 01 Process: diagCoord File: mgdArp.c Line: 154 Task ID: 0x7effdf0 Time: 1689 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Network problem detected. Check MGD subnet network configuration. arpResolve failed for IP 10.0.0.102, numTries 2, delay 100 ticks. status = -1. Or, 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Or; 2223860 VRE reset failed on ICN icn1 with MAC address XX:XX:XX:XX:XX:XX Or, 2223885 Host: hde-mr Process: logermes File: vreBmcUpCheck.pl Line: 251 Exception Class: Shutdown Function: Host: NSP System startup failed due to failure to communicate with the VRE!-> Extended Error: The system startup failure is caused by the failure to communicate with all configured ICNs. Please check that power is applied to all ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.
Error message
2223883 VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs See long description Mon Jan 26 08:47:14 2009 Error: 2223883 Host: OVMRI1 Process: NSP File: VREReset.cc Line: 937 Exception Class: Unknown Function: Recon: Data Acquisition VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs.-> Extended Error: The TPS reset failure is caused by the failure to communicate with all configured ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.