Missing software options after upgrade due to inability to restoreINFO due to HP8400 Host PC shipped without SCSI card d…
Missing software options after upgrade due to inability to restoreINFO due to HP8400 Host PC shipped without SCSI card due to End of Life.
Symptom
Connectpro Purchased Option Missing after HDxT Upgrade See Long Description The FE finished performing a Signa HDxT upgrade and then discovered that the ConnectPro option was missing. The customer had owned this option prior to the upgrade so they should have it after the upgrade also. Normally all pre-existing options are converted to the new Host ID and installed automatically during the restoreINFO during the LFC. But the FE didn't restoreINFO during the upgrade as he was told it was better to rerun all calibrations due to the large step between releases since all hardware was replaced except for the magnet. The key is not present in eLicense for this order as it was a pre-existing key. He called the OnLine Center hoping someone could generate a new key for him, but unless Sales places this option onto the order, the OnLine Center can't do this. It has to be present in eLicense first for this order before it can be generated. How does he get the key back?
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.