Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000737250359

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware stopped with WARNING: DAS Temperature is nearing the Upper Limit.

Error message

Error: 260007372 Data Acquisition System Is Over Temperature. See LD for more details Mon Aug 28 06:39:43 2006 Error: 260007372 Host: orp Process: orp_subsys File: ctemptst.c Line: 534 Mon Aug 28 06:39:41 2006 Host : Das Ermes # : 260007372 Exception Class : Pause Severity : Secondary File : ctemptst.c Line# : 534 Function : No System Function Reported Scan Type : Static Scan: 50359/2/1 Data Acquisition System Is Over Temperature.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5324279 U0522AM_FAN AXIAL,172x150x51 5324279 U0522AM_FAN AXIAL,172x150x51 Part #: 5324279 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: U0522AM_FAN AXIAL,172x150x51 AC115 FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.