Symptom

GT table would reset when cradle is driven. Up and down work without error See Long Description for more details GT table would reset when driven. Two amber led's lit on the GTCB, cradle fault and ims fault. Up and down worked.

Error message

Error: 260115183 Cradle drive fault occur see long Description Thu Mar 26 17:08:51 2009 Host : Table Sub-System Ermes # : 260115183 Exception Class : Secondary Severity : Pri/Soft File : mtcb_cmc_action.c Line# : 494 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle drive fault occur FaultCode : Resolver Fault ExtFault : Not Assigned or FaultCode : Velocity feedback(VelFB) over speed ExtFault : Variable Resolver Overspeed. |VelFB| > max(VelLmtxx)*1.5

Error: 260115184 Cradle drive fatal fault occur Cradle drive fatal fault occur FaultCode : Not Assigned ExtFault : Not Assigned Tue Mar 8 05:52:44 2011 Error: 260115184 Host: tgp Process: tgp_subsys File: mtcb_cmc_action.c Line: 526 Tue Mar 8 05:52:44 2011 Host : Table Sub-System Ermes # : 260115184 Exception Class : Secondary Severity : Pri/Soft File : mtcb_cmc_action.c Line# : 526 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle drive fatal fault occur FaultCode : Velocity feedback(VelFB) over speed ExtFault : Variable Resolver Overspeed. |VelFB| > max(VelLmtxx)*1.5

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5306599 (replaces 5132669-2 and 5132669-3) Step Motor Driver Positioning GT P - used on GT and VT tables See Long Description for details Part #: 5306599 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: Per Kittyhawk RSPL, 212008898, 6/9/08 Part #: 5132669-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Part #: 5132669-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: added per to PET Table RSPL. source pole/planner code revised per to Global Table transfer
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.